SLUUCF2C January   2021  – May 2022 BQ769142

 

  1.   Read This First
    1.     Battery Monitor Notational Conventions
    2.     Trademarks
    3.     Glossary
  2. Introduction
  3. Device Description
    1. 2.1 Overview
    2. 2.2 Functional Block Diagram
  4. Device Configuration
    1. 3.1 Direct Commands and Subcommands
    2. 3.2 Configuration Using OTP or Registers
    3. 3.3 Data Formats
      1. 3.3.1 Unsigned Integer
      2. 3.3.2 Integer
      3. 3.3.3 Floating Point
      4. 3.3.4 Hex
  5. Measurement Subsystem
    1. 4.1  Voltage Measurement
      1. 4.1.1 Voltage Measurement Schedule
      2. 4.1.2 Usage of VC Pins for Cells Versus Interconnect
      3. 4.1.3 Cell Interconnect Resistance
    2. 4.2  General Purpose ADCIN Functionality
    3. 4.3  Coulomb Counter and Digital Filters
    4. 4.4  Synchronized Voltage and Current Measurement
    5. 4.5  Subcommands 0x0071–0x0074 DASTATUS1-4(), Cell Voltage and Synchronized Current Counts
    6. 4.6  Subcommands 0x0075–0x0077 DASTATUS5–7(), Additional Measurements
    7. 4.7  Internal Temperature Measurement
    8. 4.8  Thermistor Temperature Measurement
    9. 4.9  Factory Trim of Voltage ADC
    10. 4.10 Voltage Calibration (ADC Measurements)
    11. 4.11 Voltage Calibration (COV and CUV Protections)
    12. 4.12 Current Calibration
    13. 4.13 Temperature Calibration
  6. Primary and Secondary Protection Subsystems
    1. 5.1 Protections Overview
    2. 5.2 Primary Protections
      1. 5.2.1  Primary Protections Overview
      2. 5.2.2  High-Side NFET Drivers
      3. 5.2.3  Protection FETs Configuration and Control
        1. 5.2.3.1 FET Configuration
        2. 5.2.3.2 FET Control
          1. 5.2.3.2.1 PRECHARGE Mode
          2. 5.2.3.2.2 PREDISCHARGE Mode
      4. 5.2.4  Cell Overvoltage Protection
      5. 5.2.5  Cell Undervoltage Protection
      6. 5.2.6  Short Circuit in Discharge Protection
      7. 5.2.7  Overcurrent in Charge Protection
      8. 5.2.8  Overcurrent in Discharge 1, 2, and 3 Protections
      9. 5.2.9  Overtemperature in Charge Protection
      10. 5.2.10 Overtemperature in Discharge Protection
      11. 5.2.11 Overtemperature FET Protection
      12. 5.2.12 Internal Overtemperature Protection
      13. 5.2.13 Undertemperature in Charge Protection
      14. 5.2.14 Undertemperature in Discharge Protection
      15. 5.2.15 Internal Undertemperature Protection
      16. 5.2.16 Host Watchdog Protection
      17. 5.2.17 Precharge Timeout Protection
      18. 5.2.18 Load Detect Functionality
    3. 5.3 Secondary Protections
      1. 5.3.1  Secondary Protections Overview
      2. 5.3.2  Copper Deposition (CUDEP) Permanent Fail
      3. 5.3.3  Safety Undervoltage (SUV) Permanent Fail
      4. 5.3.4  Safety Overvoltage (SOV) Permanent Fail
      5. 5.3.5  Safety Overcurrent in Charge (SOCC) Permanent Fail
      6. 5.3.6  Safety Overcurrent in Discharge (SOCD) Permanent Fail
      7. 5.3.7  Safety Cell Overtemperature (SOT) Permanent Fail
      8. 5.3.8  Safety FET Overtemperature (SOTF) Permanent Fail
      9. 5.3.9  Charge FET (CFETF) Permanent Fail
      10. 5.3.10 Discharge FET (DFETF) Permanent Fail
      11. 5.3.11 Secondary Protector (2LVL) Permanent Fail
      12. 5.3.12 Voltage Imbalance in Relax (VIMR) Permanent Fail
      13. 5.3.13 Voltage Imbalance in Active (VIMA) Permanent Fail
      14. 5.3.14 Short Circuit in Discharge Latched Permanent Fail
      15. 5.3.15 OTP Memory Signature Permanent Fail
      16. 5.3.16 Data ROM Memory Signature Permanent Fail
      17. 5.3.17 Instruction ROM Memory Signature Permanent Fail
      18. 5.3.18 LFO Oscillator Permanent Fail
      19. 5.3.19 Voltage Reference Permanent Fail
      20. 5.3.20 VSS Permanent Fail
      21. 5.3.21 Protection Comparator MUX Permanent Fail
      22. 5.3.22 Commanded Permanent Fail
      23. 5.3.23 Top of Stack Measurement Check
      24. 5.3.24 Cell Open Wire
  7. Device Status and Controls
    1. 6.1 0x00 Control Status() and 0x12 Battery Status() Commands
    2. 6.2 0x0070 MANU_DATA() Subcommand
    3. 6.3 LDOs
      1. 6.3.1 Preregulator Control
      2. 6.3.2 REG1 and REG2 LDO Controls
    4. 6.4 Multifunction Pin Controls
    5. 6.5 CFETOFF, DFETOFF, and BOTHOFF Pin Functionality
    6. 6.6 ALERT Pin Operation
    7. 6.7 DDSG and DCHG Pin Operation
    8. 6.8 Fuse Drive
    9. 6.9 Device Event Timing
  8. Operational Modes
    1. 7.1 Overview
    2. 7.2 NORMAL Mode
    3. 7.3 SLEEP Mode
    4. 7.4 DEEPSLEEP Mode
    5. 7.5 SHUTDOWN Mode
    6. 7.6 CONFIG_UPDATE Mode
  9. Device Security
    1. 8.1 Overview
  10. Serial Communications Interfaces
    1. 9.1 Serial Communications Overview
    2. 9.2 I2C Communications Subsystem
    3. 9.3 SPI Communications Interface
      1. 9.3.1 SPI Protocol
    4. 9.4 HDQ Communications Interface
  11. 10Cell Balancing
    1. 10.1 Cell Balancing Operation
    2. 10.2 Cell Balancing Timing
  12. 11Diagnostics
    1. 11.1 Diagnostics Overview
    2. 11.2 VREF2 Versus VREF1 Check
    3. 11.3 VSS Measurement
    4. 11.4 Top of Stack Measurement Check
    5. 11.5 LFO Oscillator Monitor
    6. 11.6 Protection Comparator Mux Check
    7. 11.7 Internal Watchdog Reset
    8. 11.8 Internal Memory Checks
  13. 12Commands and Subcommands
    1. 12.1 Direct Commands
    2. 12.2 Bitfield Definitions for Direct Commands
      1. 12.2.1  Control Status Register
      2. 12.2.2  Safety Alert A Register
      3. 12.2.3  Safety Status A Register
      4. 12.2.4  Safety Alert B Register
      5. 12.2.5  Safety Status B Register
      6. 12.2.6  Safety Alert C Register
      7. 12.2.7  Safety Status C Register
      8. 12.2.8  PF Alert A Register
      9. 12.2.9  PF Status A Register
      10. 12.2.10 PF Alert B Register
      11. 12.2.11 PF Status B Register
      12. 12.2.12 PF Alert C Register
      13. 12.2.13 PF Status C Register
      14. 12.2.14 PF Alert D Register
      15. 12.2.15 PF Status D Register
      16. 12.2.16 Battery Status Register
      17. 12.2.17 Alarm Status Register
      18. 12.2.18 Alarm Raw Status Register
      19. 12.2.19 Alarm Enable Register
      20. 12.2.20 FET Status Register
    3. 12.3 Command-Only Subcommands
    4. 12.4 Subcommands with Data
    5. 12.5 Bitfield Definitions for Subcommands
      1. 12.5.1 PF Status A Register
      2. 12.5.2 PF Status B Register
      3. 12.5.3 PF Status C Register
      4. 12.5.4 PF Status D Register
      5. 12.5.5 Manufacturing Status Register
      6. 12.5.6 FET Control Register
      7. 12.5.7 REG12 Control Register
      8. 12.5.8 OTP Write Check Result Register
      9. 12.5.9 OTP Write Result Register
  14. 13Data Memory Settings
    1. 13.1 Data Memory Access
    2. 13.2 Calibration
      1. 13.2.1  Calibration:Voltage
        1. 13.2.1.1  Calibration:Voltage:Cell 1 Gain
        2. 13.2.1.2  Calibration:Voltage:Cell 2 Gain
        3. 13.2.1.3  Calibration:Voltage:Cell 3 Gain
        4. 13.2.1.4  Calibration:Voltage:Cell 4 Gain
        5. 13.2.1.5  Calibration:Voltage:Cell 5 Gain
        6. 13.2.1.6  Calibration:Voltage:Cell 6 Gain
        7. 13.2.1.7  Calibration:Voltage:Cell 7 Gain
        8. 13.2.1.8  Calibration:Voltage:Cell 8 Gain
        9. 13.2.1.9  Calibration:Voltage:Cell 9 Gain
        10. 13.2.1.10 Calibration:Voltage:Cell 10 Gain
        11. 13.2.1.11 Calibration:Voltage:Cell 11 Gain
        12. 13.2.1.12 Calibration:Voltage:Cell 12 Gain
        13. 13.2.1.13 Calibration:Voltage:Cell 13 Gain
        14. 13.2.1.14 Calibration:Voltage:Cell 14 Gain
        15. 13.2.1.15 Calibration:Voltage:Pack Gain
        16. 13.2.1.16 Calibration:Voltage:TOS Gain
        17. 13.2.1.17 Calibration:Voltage:LD Gain
        18. 13.2.1.18 Calibration:Voltage:ADC Gain
      2. 13.2.2  Calibration:Current
        1. 13.2.2.1 Calibration:Current:CC Gain
        2. 13.2.2.2 Calibration:Current:Capacity Gain
      3. 13.2.3  Calibration:Vcell Offset
        1. 13.2.3.1 Calibration:Vcell Offset:Vcell Offset
      4. 13.2.4  Calibration:V Divider Offset
        1. 13.2.4.1 Calibration:V Divider Offset:Vdiv Offset
      5. 13.2.5  Calibration:Current Offset
        1. 13.2.5.1 Calibration:Current Offset:Coulomb Counter Offset Samples
        2. 13.2.5.2 Calibration:Current Offset:Board Offset
      6. 13.2.6  Calibration:Temperature
        1. 13.2.6.1  Calibration:Temperature:Internal Temp Offset
        2. 13.2.6.2  Calibration:Temperature:CFETOFF Temp Offset
        3. 13.2.6.3  Calibration:Temperature:DFETOFF Temp Offset
        4. 13.2.6.4  Calibration:Temperature:ALERT Temp Offset
        5. 13.2.6.5  Calibration:Temperature:TS1 Temp Offset
        6. 13.2.6.6  Calibration:Temperature:TS2 Temp Offset
        7. 13.2.6.7  Calibration:Temperature:TS3 Temp Offset
        8. 13.2.6.8  Calibration:Temperature:HDQ Temp Offset
        9. 13.2.6.9  Calibration:Temperature:DCHG Temp Offset
        10. 13.2.6.10 Calibration:Temperature:DDSG Temp Offset
      7. 13.2.7  Calibration:Internal Temp Model
        1. 13.2.7.1 Calibration:Internal Temp Model:Int Gain
        2. 13.2.7.2 Calibration:Internal Temp Model:Int base offset
        3. 13.2.7.3 Calibration:Internal Temp Model:Int Maximum AD
        4. 13.2.7.4 Calibration:Internal Temp Model:Int Maximum Temp
      8. 13.2.8  Calibration:18K Temperature Model
        1. 13.2.8.1  Calibration:18K Temperature Model:Coeff a1
        2. 13.2.8.2  Calibration:18K Temperature Model:Coeff a2
        3. 13.2.8.3  Calibration:18K Temperature Model:Coeff a3
        4. 13.2.8.4  Calibration:18K Temperature Model:Coeff a4
        5. 13.2.8.5  Calibration:18K Temperature Model:Coeff a5
        6. 13.2.8.6  Calibration:18K Temperature Model:Coeff b1
        7. 13.2.8.7  Calibration:18K Temperature Model:Coeff b2
        8. 13.2.8.8  Calibration:18K Temperature Model:Coeff b3
        9. 13.2.8.9  Calibration:18K Temperature Model:Coeff b4
        10. 13.2.8.10 Calibration:18K Temperature Model:Adc0
      9. 13.2.9  Calibration:180K Temperature Model
        1. 13.2.9.1  Calibration:180K Temperature Model:Coeff a1
        2. 13.2.9.2  Calibration:180K Temperature Model:Coeff a2
        3. 13.2.9.3  Calibration:180K Temperature Model:Coeff a3
        4. 13.2.9.4  Calibration:180K Temperature Model:Coeff a4
        5. 13.2.9.5  Calibration:180K Temperature Model:Coeff a5
        6. 13.2.9.6  Calibration:180K Temperature Model:Coeff b1
        7. 13.2.9.7  Calibration:180K Temperature Model:Coeff b2
        8. 13.2.9.8  Calibration:180K Temperature Model:Coeff b3
        9. 13.2.9.9  Calibration:180K Temperature Model:Coeff b4
        10. 13.2.9.10 Calibration:180K Temperature Model:Adc0
      10. 13.2.10 Calibration:Custom Temperature Model
        1. 13.2.10.1  Calibration:Custom Temperature Model:Coeff a1
        2. 13.2.10.2  Calibration:Custom Temperature Model:Coeff a2
        3. 13.2.10.3  Calibration:Custom Temperature Model:Coeff a3
        4. 13.2.10.4  Calibration:Custom Temperature Model:Coeff a4
        5. 13.2.10.5  Calibration:Custom Temperature Model:Coeff a5
        6. 13.2.10.6  Calibration:Custom Temperature Model:Coeff b1
        7. 13.2.10.7  Calibration:Custom Temperature Model:Coeff b2
        8. 13.2.10.8  Calibration:Custom Temperature Model:Coeff b3
        9. 13.2.10.9  Calibration:Custom Temperature Model:Coeff b4
        10. 13.2.10.10 Calibration:Custom Temperature Model:Rc0
        11. 13.2.10.11 Calibration:Custom Temperature Model:Adc0
      11. 13.2.11 Calibration:Current Deadband
        1. 13.2.11.1 Calibration:Current Deadband:Coulomb Counter Deadband
      12. 13.2.12 Calibration:CUV
        1. 13.2.12.1 Calibration:CUV:CUV Threshold Override
      13. 13.2.13 Calibration:COV
        1. 13.2.13.1 Calibration:COV:COV Threshold Override
    3. 13.3 Settings
      1. 13.3.1  Settings:Fuse
        1. 13.3.1.1 Settings:Fuse:Min Blow Fuse Voltage
        2. 13.3.1.2 Settings:Fuse:Fuse Blow Timeout
      2. 13.3.2  Settings:Configuration
        1. 13.3.2.1  Settings:Configuration:Power Config
        2. 13.3.2.2  Settings:Configuration:REG12 Config
        3. 13.3.2.3  Settings:Configuration:REG0 Config
        4. 13.3.2.4  Settings:Configuration:HWD Regulator Options
        5. 13.3.2.5  Settings:Configuration:Comm Type
        6. 13.3.2.6  Settings:Configuration:I2C Address
        7. 13.3.2.7  Settings:Configuration:SPI Configuration
        8. 13.3.2.8  Settings:Configuration:Comm Idle Time
        9. 13.3.2.9  Settings:Configuration:CFETOFF Pin Config
        10. 13.3.2.10 Settings:Configuration:DFETOFF Pin Config
        11. 13.3.2.11 Settings:Configuration:ALERT Pin Config
        12. 13.3.2.12 Settings:Configuration:TS1 Config
        13. 13.3.2.13 Settings:Configuration:TS2 Config
        14. 13.3.2.14 Settings:Configuration:TS3 Config
        15. 13.3.2.15 Settings:Configuration:HDQ Pin Config
        16. 13.3.2.16 Settings:Configuration:DCHG Pin Config
        17. 13.3.2.17 Settings:Configuration:DDSG Pin Config
        18. 13.3.2.18 Settings:Configuration:DA Configuration
        19. 13.3.2.19 Settings:Configuration:Vcell Mode
        20. 13.3.2.20 Settings:Configuration:CC3 Samples
      3. 13.3.3  Settings:Protection
        1. 13.3.3.1  Settings:Protection:Protection Configuration
        2. 13.3.3.2  Settings:Protection:Enabled Protections A
        3. 13.3.3.3  Settings:Protection:Enabled Protections B
        4. 13.3.3.4  Settings:Protection:Enabled Protections C
        5. 13.3.3.5  Settings:Protection:CHG FET Protections A
        6. 13.3.3.6  Settings:Protection:CHG FET Protections B
        7. 13.3.3.7  Settings:Protection:CHG FET Protections C
        8. 13.3.3.8  Settings:Protection:DSG FET Protections A
        9. 13.3.3.9  Settings:Protection:DSG FET Protections B
        10. 13.3.3.10 Settings:Protection:DSG FET Protections C
        11. 13.3.3.11 Settings:Protection:Body Diode Threshold
      4. 13.3.4  Settings:Alarm
        1. 13.3.4.1 Settings:Alarm:Default Alarm Mask
        2. 13.3.4.2 Settings:Alarm:SF Alert Mask A
        3. 13.3.4.3 Settings:Alarm:SF Alert Mask B
        4. 13.3.4.4 Settings:Alarm:SF Alert Mask C
        5. 13.3.4.5 Settings:Alarm:PF Alert Mask A
        6. 13.3.4.6 Settings:Alarm:PF Alert Mask B
        7. 13.3.4.7 Settings:Alarm:PF Alert Mask C
        8. 13.3.4.8 Settings:Alarm:PF Alert Mask D
      5. 13.3.5  Settings:Permanent Failure
        1. 13.3.5.1 Settings:Permanent Failure:Enabled PF A
        2. 13.3.5.2 Settings:Permanent Failure:Enabled PF B
        3. 13.3.5.3 Settings:Permanent Failure:Enabled PF C
        4. 13.3.5.4 Settings:Permanent Failure:Enabled PF D
      6. 13.3.6  Settings:FET
        1. 13.3.6.1 Settings:FET:FET Options
        2. 13.3.6.2 Settings:FET:Chg Pump Control
        3. 13.3.6.3 Settings:FET:Precharge Start Voltage
        4. 13.3.6.4 Settings:FET:Precharge Stop Voltage
        5. 13.3.6.5 Settings:FET:Predischarge Timeout
        6. 13.3.6.6 Settings:FET:Predischarge Stop Delta
      7. 13.3.7  Settings:Current Thresholds
        1. 13.3.7.1 Settings:Current Thresholds:Dsg Current Threshold
        2. 13.3.7.2 Settings:Current Thresholds:Chg Current Threshold
      8. 13.3.8  Settings:Cell Open-Wire
        1. 13.3.8.1 Settings:Cell Open-Wire:Check Time
      9. 13.3.9  Settings:Interconnect Resistances
        1. 13.3.9.1  Settings:Interconnect Resistances:Cell 1 Interconnect
        2. 13.3.9.2  Settings:Interconnect Resistances:Cell 2 Interconnect
        3. 13.3.9.3  Settings:Interconnect Resistances:Cell 3 Interconnect
        4. 13.3.9.4  Settings:Interconnect Resistances:Cell 4 Interconnect
        5. 13.3.9.5  Settings:Interconnect Resistances:Cell 5 Interconnect
        6. 13.3.9.6  Settings:Interconnect Resistances:Cell 6 Interconnect
        7. 13.3.9.7  Settings:Interconnect Resistances:Cell 7 Interconnect
        8. 13.3.9.8  Settings:Interconnect Resistances:Cell 8 Interconnect
        9. 13.3.9.9  Settings:Interconnect Resistances:Cell 9 Interconnect
        10. 13.3.9.10 Settings:Interconnect Resistances:Cell 10 Interconnect
        11. 13.3.9.11 Settings:Interconnect Resistances:Cell 11 Interconnect
        12. 13.3.9.12 Settings:Interconnect Resistances:Cell 12 Interconnect
        13. 13.3.9.13 Settings:Interconnect Resistances:Cell 13 Interconnect
        14. 13.3.9.14 Settings:Interconnect Resistances:Cell 14 Interconnect
      10. 13.3.10 Settings:Manufacturing
        1. 13.3.10.1 Settings:Manufacturing:Mfg Status Init
      11. 13.3.11 Settings:Cell Balancing Config
        1. 13.3.11.1  Settings:Cell Balancing Config:Balancing Configuration
        2. 13.3.11.2  Settings:Cell Balancing Config:Min Cell Temp
        3. 13.3.11.3  Settings:Cell Balancing Config:Max Cell Temp
        4. 13.3.11.4  Settings:Cell Balancing Config:Max Internal Temp
        5. 13.3.11.5  Settings:Cell Balancing Config:Cell Balance Interval
        6. 13.3.11.6  Settings:Cell Balancing Config:Cell Balance Max Cells
        7. 13.3.11.7  Settings:Cell Balancing Config:Cell Balance Min Cell V (Charge)
        8. 13.3.11.8  Settings:Cell Balancing Config:Cell Balance Min Delta (Charge)
        9. 13.3.11.9  Settings:Cell Balancing Config:Cell Balance Stop Delta (Charge)
        10. 13.3.11.10 Settings:Cell Balancing Config:Cell Balance Min Cell V (Relax)
        11. 13.3.11.11 Settings:Cell Balancing Config:Cell Balance Min Delta (Relax)
        12. 13.3.11.12 Settings:Cell Balancing Config:Cell Balance Stop Delta (Relax)
    4. 13.4 Power
      1. 13.4.1 Power:Shutdown
        1. 13.4.1.1 Power:Shutdown:Shutdown Cell Voltage
        2. 13.4.1.2 Power:Shutdown:Shutdown Stack Voltage
        3. 13.4.1.3 Power:Shutdown:Low V Shutdown Delay
        4. 13.4.1.4 Power:Shutdown:Shutdown Temperature
        5. 13.4.1.5 Power:Shutdown:Shutdown Temperature Delay
        6. 13.4.1.6 Power:Shutdown:FET Off Delay
        7. 13.4.1.7 Power:Shutdown:Shutdown Command Delay
        8. 13.4.1.8 Power:Shutdown:Auto Shutdown Time
        9. 13.4.1.9 Power:Shutdown:RAM Fail Shutdown Time
      2. 13.4.2 Power:Sleep
        1. 13.4.2.1 Power:Sleep:Sleep Current
        2. 13.4.2.2 Power:Sleep:Voltage Time
        3. 13.4.2.3 Power:Sleep:Wake Comparator Current
        4. 13.4.2.4 Power:Sleep:Sleep Hysteresis Time
        5. 13.4.2.5 Power:Sleep:Sleep Charger Voltage Threshold
        6. 13.4.2.6 Power:Sleep:Sleep Charger PACK-TOS Delta
    5. 13.5 System Data
      1. 13.5.1 System Data:Integrity
        1. 13.5.1.1 System Data:Integrity:Config RAM Signature
    6. 13.6 Protections
      1. 13.6.1  Protections:CUV
        1. 13.6.1.1 Protections:CUV:Threshold
        2. 13.6.1.2 Protections:CUV:Delay
        3. 13.6.1.3 Protections:CUV:Recovery Hysteresis
      2. 13.6.2  Protections:COV
        1. 13.6.2.1 Protections:COV:Threshold
        2. 13.6.2.2 Protections:COV:Delay
        3. 13.6.2.3 Protections:COV:Recovery Hysteresis
      3. 13.6.3  Protections:COVL
        1. 13.6.3.1 Protections:COVL:Latch Limit
        2. 13.6.3.2 Protections:COVL:Counter Dec Delay
        3. 13.6.3.3 Protections:COVL:Recovery Time
      4. 13.6.4  Protections:OCC
        1. 13.6.4.1 Protections:OCC:Threshold
        2. 13.6.4.2 Protections:OCC:Delay
        3. 13.6.4.3 Protections:OCC:Recovery Threshold
        4. 13.6.4.4 Protections:OCC:PACK-TOS Delta
      5. 13.6.5  Protections:OCD1
        1. 13.6.5.1 Protections:OCD1:Threshold
        2. 13.6.5.2 Protections:OCD1:Delay
      6. 13.6.6  Protections:OCD2
        1. 13.6.6.1 Protections:OCD2:Threshold
        2. 13.6.6.2 Protections:OCD2:Delay
      7. 13.6.7  Protections:SCD
        1. 13.6.7.1 Protections:SCD:Threshold
        2. 13.6.7.2 Protections:SCD:Delay
        3. 13.6.7.3 Protections:SCD:Recovery Time
      8. 13.6.8  Protections:OCD3
        1. 13.6.8.1 Protections:OCD3:Threshold
        2. 13.6.8.2 Protections:OCD3:Delay
      9. 13.6.9  Protections:OCD
        1. 13.6.9.1 Protections:OCD:Recovery Threshold
      10. 13.6.10 Protections:OCDL
        1. 13.6.10.1 Protections:OCDL:Latch Limit
        2. 13.6.10.2 Protections:OCDL:Counter Dec Delay
        3. 13.6.10.3 Protections:OCDL:Recovery Time
        4. 13.6.10.4 Protections:OCDL:Recovery Threshold
      11. 13.6.11 Protections:SCDL
        1. 13.6.11.1 Protections:SCDL:Latch Limit
        2. 13.6.11.2 Protections:SCDL:Counter Dec Delay
        3. 13.6.11.3 Protections:SCDL:Recovery Time
        4. 13.6.11.4 Protections:SCDL:Recovery Threshold
      12. 13.6.12 Protections:OTC
        1. 13.6.12.1 Protections:OTC:Threshold
        2. 13.6.12.2 Protections:OTC:Delay
        3. 13.6.12.3 Protections:OTC:Recovery
      13. 13.6.13 Protections:OTD
        1. 13.6.13.1 Protections:OTD:Threshold
        2. 13.6.13.2 Protections:OTD:Delay
        3. 13.6.13.3 Protections:OTD:Recovery
      14. 13.6.14 Protections:OTF
        1. 13.6.14.1 Protections:OTF:Threshold
        2. 13.6.14.2 Protections:OTF:Delay
        3. 13.6.14.3 Protections:OTF:Recovery
      15. 13.6.15 Protections:OTINT
        1. 13.6.15.1 Protections:OTINT:Threshold
        2. 13.6.15.2 Protections:OTINT:Delay
        3. 13.6.15.3 Protections:OTINT:Recovery
      16. 13.6.16 Protections:UTC
        1. 13.6.16.1 Protections:UTC:Threshold
        2. 13.6.16.2 Protections:UTC:Delay
        3. 13.6.16.3 Protections:UTC:Recovery
      17. 13.6.17 Protections:UTD
        1. 13.6.17.1 Protections:UTD:Threshold
        2. 13.6.17.2 Protections:UTD:Delay
        3. 13.6.17.3 Protections:UTD:Recovery
      18. 13.6.18 Protections:UTINT
        1. 13.6.18.1 Protections:UTINT:Threshold
        2. 13.6.18.2 Protections:UTINT:Delay
        3. 13.6.18.3 Protections:UTINT:Recovery
      19. 13.6.19 Protections:Recovery
        1. 13.6.19.1 Protections:Recovery:Time
      20. 13.6.20 Protections:HWD
        1. 13.6.20.1 Protections:HWD:Delay
      21. 13.6.21 Protections:Load Detect
        1. 13.6.21.1 Protections:Load Detect:Active Time
        2. 13.6.21.2 Protections:Load Detect:Retry Delay
        3. 13.6.21.3 Protections:Load Detect:Timeout
      22. 13.6.22 Protections:PTO
        1. 13.6.22.1 Protections:PTO:Charge Threshold
        2. 13.6.22.2 Protections:PTO:Delay
        3. 13.6.22.3 Protections:PTO:Reset
    7. 13.7 Permanent Fail
      1. 13.7.1  Permanent Fail:CUDEP
        1. 13.7.1.1 Permanent Fail:CUDEP:Threshold
        2. 13.7.1.2 Permanent Fail:CUDEP:Delay
      2. 13.7.2  Permanent Fail:SUV
        1. 13.7.2.1 Permanent Fail:SUV:Threshold
        2. 13.7.2.2 Permanent Fail:SUV:Delay
      3. 13.7.3  Permanent Fail:SOV
        1. 13.7.3.1 Permanent Fail:SOV:Threshold
        2. 13.7.3.2 Permanent Fail:SOV:Delay
      4. 13.7.4  Permanent Fail:TOS
        1. 13.7.4.1 Permanent Fail:TOS:Threshold
        2. 13.7.4.2 Permanent Fail:TOS:Delay
      5. 13.7.5  Permanent Fail:SOCC
        1. 13.7.5.1 Permanent Fail:SOCC:Threshold
        2. 13.7.5.2 Permanent Fail:SOCC:Delay
      6. 13.7.6  Permanent Fail:SOCD
        1. 13.7.6.1 Permanent Fail:SOCD:Threshold
        2. 13.7.6.2 Permanent Fail:SOCD:Delay
      7. 13.7.7  Permanent Fail:SOT
        1. 13.7.7.1 Permanent Fail:SOT:Threshold
        2. 13.7.7.2 Permanent Fail:SOT:Delay
      8. 13.7.8  Permanent Fail:SOTF
        1. 13.7.8.1 Permanent Fail:SOTF:Threshold
        2. 13.7.8.2 Permanent Fail:SOTF:Delay
      9. 13.7.9  Permanent Fail:VIMR
        1. 13.7.9.1 Permanent Fail:VIMR:Check Voltage
        2. 13.7.9.2 Permanent Fail:VIMR:Max Relax Current
        3. 13.7.9.3 Permanent Fail:VIMR:Threshold
        4. 13.7.9.4 Permanent Fail:VIMR:Delay
        5. 13.7.9.5 Permanent Fail:VIMR:Relax Min Duration
      10. 13.7.10 Permanent Fail:VIMA
        1. 13.7.10.1 Permanent Fail:VIMA:Check Voltage
        2. 13.7.10.2 Permanent Fail:VIMA:Min Active Current
        3. 13.7.10.3 Permanent Fail:VIMA:Threshold
        4. 13.7.10.4 Permanent Fail:VIMA:Delay
      11. 13.7.11 Permanent Fail:CFETF
        1. 13.7.11.1 Permanent Fail:CFETF:OFF Threshold
        2. 13.7.11.2 Permanent Fail:CFETF:OFF Delay
      12. 13.7.12 Permanent Fail:DFETF
        1. 13.7.12.1 Permanent Fail:DFETF:OFF Threshold
        2. 13.7.12.2 Permanent Fail:DFETF:OFF Delay
      13. 13.7.13 Permanent Fail:VSSF
        1. 13.7.13.1 Permanent Fail:VSSF:Fail Threshold
        2. 13.7.13.2 Permanent Fail:VSSF:Delay
      14. 13.7.14 Permanent Fail:2LVL
        1. 13.7.14.1 Permanent Fail:2LVL:Delay
      15. 13.7.15 Permanent Fail:LFOF
        1. 13.7.15.1 Permanent Fail:LFOF:Delay
      16. 13.7.16 Permanent Fail:HWMX
        1. 13.7.16.1 Permanent Fail:HWMX:Delay
    8. 13.8 Security
      1. 13.8.1 Security:Settings
        1. 13.8.1.1 Security:Settings:Security Settings
      2. 13.8.2 Security:Keys
        1. 13.8.2.1 Security:Keys:Unseal Key Step 1
        2. 13.8.2.2 Security:Keys:Unseal Key Step 2
        3. 13.8.2.3 Security:Keys:Full Access Key Step 1
        4. 13.8.2.4 Security:Keys:Full Access Key Step 2
    9. 13.9 Data Memory Summary
  15. 15Revision History

Settings:Configuration:Power Config

Class Subclass Name Type Min Max Default Unit
Settings Configuration Power Config H2 0x0000 0xFFFF 0x2982 Hex
15 14 13 12 11 10 9 8
RSVD_0 RSVD_0 DPSLP_OT SHUT_TS2 DPSLP_PD DPSLP_LDO DPSLP_LFO SLEEP
7 6 5 4 3 2 1 0
OTSD FASTADC CB_LOOP_SLOW_1 CB_LOOP_SLOW_0 LOOP_SLOW_1 LOOP_SLOW_0 WK_SPD_1 WK_SPD_0
Table 13-1 Power Config Register Field Descriptions
Bit Field Default Description
13 DPSLP_OT 1 Enables transition from DEEPSLEEP to SHUTDOWN based on on-chip overtemperature detection.

0 = In DEEPSLEEP, on-chip OT is disabled

1 = On-chip OT is enabled in DEEPSLEEP, allowing transition to SHUTDOWN

12 SHUT_TS2 0 When TS2 is used as a thermistor or has some other external pulldown, SHUTDOWN mode cannot be used. If TS2 is externally pulled down, this bit should be set to prevent the device from immediately exiting SHUTDOWN mode in some circumstances. When this bit is set and SHUTDOWN conditions are met, the device will instead enter a low power state waiting for a rising edge on the LD pin.

0 = Standard SHUTDOWN mode behavior is selected.

1 = SHUTDOWN mode replaced by low-power state waiting for rising edge on LD pin

11 DPSLP_PD 1 Enables wake from DEEPSLEEP based on charger attach. When clear, DEEPSLEEP mode must be exited via host command.

0 = DEEPSLEEP not exited on rising edge on LD pin

1 = A rising edge on the LD pin exits DEEPSLEEP

10 DPSLP_LDO 0 Determines whether or not REG1 and REG2 are disabled in DEEPSLEEP mode

0 = Disables REG1 and REG2 when entering DEEPSLEEP mode

1 = Leaves REG1 and REG2 in present state when entering DEEPSLEEP

9 DPSLP_LFO 0 Determines whether or not to disable the Low Frequency Oscillator in DEEPSLEEP mode to conserve power.

0 = Disables the Low Frequency Oscillator in DEEPSLEEP mode (recommended)

1 = Enables the Low Frequency Oscillator in DEEPSLEEP mode

8 SLEEP 1 Sets the default value of BatteryStatus()[SLEEP_EN] which enables or disables SLEEP mode. After initialization, SLEEP_EN can stil be changed via the SLEEP_ENABLE and SLEEP_DISABLE subcommands.

0 = Disables SLEEP mode by default

1 = Enables SLEEP mode by default

7 OTSD 1 Enables or disables the on-chip overtemperature detection circuit to shut down the device in case of a severe on-chip overtemperature condition.

0 = Disables SHUTDOWN from on-chip overtemperature detection circuit (not recommended)

1 = Enters SHUTDOWN mode when an on-chip overtemperature condition is detected

6 FASTADC 0 Selects ADC conversion speed for voltage and simultaneous current measurements. Higher speed results in lower accuracy.

0 = 3 ms per conversion

1 = 1.5 ms per conversion

5–4 CB_LOOP_SLOW_1–CB_LOOP_SLOW_0 0 Selects ADC scan loop speed while cell balancing is active by inserting current-only measurements after each voltage and temperature scan loop. This can be used to slow down voltage measurements while balancing to increase the duty-cycle, since balancing must be paused during measurement of the cell.

0 = Full speed

1 = Half speed

2 = Quarter speed

3 = Eighth speed

3–2 LOOP_SLOW_1–LOOP_SLOW_0 0 Selects normal ADC scan loop speed by inserting current-only measurements after each voltage and temperature scan loop. This setting is only used while cell balancing is not active.

0 = Full speed

1 = Half speed

2 = Quarter speed

3 = Eighth speed

1–0 WK_SPD_1–WK_SPD_0 2

Selects the coulomb counter conversion speed used in SLEEP mode for the current wake comparator function. Slower conversion speed results in lower noise. Setting 0x2 may exhibit a large offset and should be avoided. Setting 0x3 results in ~100μV (1-sigma) noise level, so should only be used when the Wake Comparator Current is set to a level such that |VSRP–VSRN| > 1000 μV.

0x0 = 48 ms conversion rate

0x1 = 24 ms conversion rate

0x2 = 12 ms conversion rate

0x3 = 6 ms conversion rate