AFE4490
- Fully-Integrated Analog Front-End for
Pulse Oximeter Applications:- Flexible Pulse Sequencing and
Timing Control
- Flexible Pulse Sequencing and
- Transmit:
- Integrated LED Driver (H-Bridge, Push, or Pull)
- 110-dB Dynamic Range Across Full Range (Enables Low Noise at Low LED Current)
- LED Current:
- Programmable Ranges of 50 mA, 75 mA, 100 mA, 150 mA, and 200 mA,
Each with 8-Bit Current Resolution
- Programmable Ranges of 50 mA, 75 mA, 100 mA, 150 mA, and 200 mA,
- Low Power:
- 100 µA + Average LED Current
- LED On-Time Programmability from
(50 µs + Settle Time) to 4 ms - Independent LED2, LED1 Current Reference
- Receive Channel with High Dynamic Range:
- Input-Referred Noise:
50 pARMS (at 5-µA PD Current) - 13.5 Noise-Free Bits (at 5-µA PD Current)
- Analog Ambient Cancellation Scheme with Selectable 1-µA to 10-µA Ambient Current
- Low Power: < 2.3 mW at 3.0-V Supply
- Rx Sample Time: 50 µs to 4 ms
- I-V Amplifier with Seven Separate LED2 and LED1 Programmable Feedback R and C Settings
- Integrated Digital Ambient Estimation and Subtraction
- Input-Referred Noise:
- Integrated Fault Diagnostics:
- Photodiode and LED Open and
Short Detection - Cable On or Off Detection
- Photodiode and LED Open and
- Supplies:
- Rx = 2.0 V to 3.6 V
- Tx = 3.0 V or 5.25 V
- Package: Compact VQFN-40 (6 mm × 6 mm)
- Specified Temperature Range: –40°C to 85°C
The AFE4490 is a fully-integrated analog front-end (AFE) that is ideally suited for pulse-oximeter applications. The device consists of a low-noise receiver channel with a 22-bit analog-to-digital converter (ADC), an LED transmit section, and diagnostics for sensor and LED fault detection. The device is a very configurable timing controller. This flexibility enables the user to have complete control of the device timing characteristics. To ease clocking requirements and provide a low-jitter clock to the device, an oscillator is also integrated that functions from an external crystal. The device communicates to an external microcontroller or host processor using an SPI™ interface.
The device is a complete AFE solution packaged in a single, compact VQFN-40 package (6 mm × 6 mm) and is specified over the operating temperature range of –40°C to 85°C.
技術資料
種類 | タイトル | 最新の英語版をダウンロード | 日付 | |||
---|---|---|---|---|---|---|
* | データシート | AFE4490 Integrated Analog Front-End for Pulse Oximeters データシート (Rev. H) | PDF | HTML | 2014年 10月 29日 | ||
EVM ユーザー ガイド (英語) | AFE44x0SPO2EVM User's Guide (Rev. C) | 2014年 5月 27日 |
設計および開発
その他のアイテムや必要なリソースを参照するには、以下のタイトルをクリックして詳細ページをご覧ください。
AFE4490SPO2EVM — AFE4490 パルス・オキシメータ (血中酸素飽和度計) 向け統合型アナログ・フロント・エンド (AFE) の評価基板
The AFE44x0SPO2EVM is intended for evaluating AFE4400 and AFE4490 devices. The family of devices are fully integrated AFE ideally suited for Pulse Oximeter applications and consist of low noise receive channel, the LED transmit section and diagnostics for sensor and LED fault detection. The (...)
SLAC556 — AFE44x0SPO2EVM GUI v2.4
サポート対象の製品とハードウェア
製品
バイオセンシング AFE
ハードウェア開発
評価ボード
PSPICE-FOR-TI — TI Design / シミュレーション・ツール向け PSpice®
設計とシミュレーション向けの環境である PSpice for TI (...)
パッケージ | ピン数 | CAD シンボル、フットプリント、および 3D モデル |
---|---|---|
VQFN (RHA) | 40 | Ultra Librarian |
購入と品質
- RoHS
- REACH
- デバイスのマーキング
- リード端子の仕上げ / ボールの原材料
- MSL 定格 / ピーク リフロー
- MTBF/FIT 推定値
- 使用原材料
- 認定試験結果
- 継続的な信頼性モニタ試験結果
- ファブの拠点
- 組み立てを実施した拠点
推奨製品には、この TI 製品に関連するパラメータ、評価基板、またはリファレンス デザインが存在する可能性があります。